Login Login
  •  
MORE

WIDGETS

Widgets

Wanted articles
Who is online?
Article tools

Knee joint

From Orthoload Wiki

Jump to: navigation, search


Instrumented implant

Implant: knee joint
Implant: knee joint

In order to obtain realistic loading data, a knee implant with a 9-channel telemetry transmitter was developed which enables six-component load measurements in a primary total knee replacement. Both forces in axial, medio-lateral and anterio-posterior direction and flexion-extension, varus-valgus and internal-external moments can be measured.The instrumented knee joint is a modification of the INNEXTM System, Type FIXUC (Zimmer GmbH, Winterthur, Switzerland). The standard femur component and tibial insert are used. Only the tibial component was modified to enable the integration of the electronic devices. During modification of the tibial component, the patients' safety was deemed to be especially important.

Coordinate system

Coordinate system Knee Joint
Coordinate system knee joint

The coordinate system of the instrumented knee implant is a right-handed coordinate system fixed at the right tibial implant (not at the bone!). If forces and moments are measured in a left knee, they are transformed to the right side. The coordinate system is located at the height of the lowest part of the polyethylene insert. The z-axis is aligned with the stem axis of the implant.

The force components +Fx, +Fy and +Fz act in lateral, anterior and superior direction on the tibial tray. The moment Mx acts in the sagittal plane of the tibial component and turns clockwise around the +x-axis. The moment My acts in the frontal plane and turns clockwise around the +y-axis and the moment Mz turns clockwise around +z-axis in the transverse plane. A positive moment Mz acts if the tibial implant component (or the femur) rotates inwards and/or if the tibia bone rotates outwards. The OrthoLoad videos show the load components relative to the tibial tray. The stem axis z of the tibial implant component is rotated backwards in the sagittal plane by about 7 degree relative to the long axis of the tibia bone. This slope of the implant varies inter-individually.

Patients

Pat K1L big.png Pat K2L big.png Pat K3R big.png Pat K4R big.png Pat K5R big.png
K1L K2L K3R K4R K5R
Pat K6L big.png Pat K7L big.png Pat K8L big.png Pat K9L big.png
K6L K7L K8L K9L

Table with basic information about the knee joint patients:

Patient Side Gender Weight
[kg]
Height
[cm]
Age at
Implantation
[years]
Indication
K1L left m 100 177 63 Osteoarthritis
K2L left m 93 171 71 Osteoarthritis
K3R right m 95 175 70 Osteoarthritis
K4R right f 92 170 63 Osteoarthritis
K5R right m 94 175 60 Osteoarthritis
K6L left f 76 174 65 Osteoarthritis
K7L left f 70 166 74 Osteoarthritis
K8L left m 77 174 70 Osteoarthritis
K9L left m 100 166 75 Osteoarthritis